Common Failures of Salt Spray Environmental Chambers
When you encounter some common failures of salt spray environmental chambers, how to solve them? Wuxi Guanya Refrigeration Technology Co., Ltd. has brought the relevant fault tips as follows.
Salt spray environmental chamber cannot spray
Failure causes: 1. air compressor does not work; 2. main switch of the air compressor outlet does not open; 3. solenoid valve malfunctions; 4. pressure meter malfunctions or the pressure is too low; 5. electromagnetic contactor malfunctions; 6. nozzle blocks;
Handling methods: 1. open the air compressor; 2. open the main switch of the air compressor; 3.4.5. ask the after-sales department for door-to-door repair service; 6. remove the nozzle and clean it (please carefully disassemble).
The temperature of the saturation tank of the salt spray environmental chamber cannot rise to the set temperature
Failure causes: 1. the temperature of the temperature controller of the saturation tank is too low; 2. the temperature of the saturation bucket safety protection switch is set too low; 3. heating system malfunctions; 4. electromagnetic relay malfunctions; 5. controller malfunctions.
Handling methods: 1. Set the temperature of the temperature controller to the required temperature; 2. Set the temperature of the safety switch to the required temperature; 3. Ask the after-sales department for door-to-door repair service.
Lack of spray amount in the salt spray environmental chamber
Failure causes: 1. spray regulator is placed too low; 2. glass filter pf the preheating tank is blocked; 3. pressure is set too low.
Handling methods: 1. raise the spray regulator; 2. clean up the glass filter; 3. adjust the pressure of the pressure valve to 1Kg/cm², the air compressor is marked with a pressure valve, adjust to the pressure to 2Kg/cm².
These are some handling methods of the salt spray environmental chamber failures brought by Wuxi LNEYA. Wuxi LNEYA has environmental chambers for sale, the quality and the after-sales are guaranteed.
Raccomandazioni correlate
-
Descrizione delle prestazioni della camera di prova per gli shock termici
1901La camera di prova a shock termico è ampiamente utilizzata in diversi settori industriali importanti. Al momento dell'acquisto, gli utenti sperano che la loro camera di prova per urti termici possa essere utilizzata a lungo. Pertanto, è necessario comprendere ogni difetto della camera...
Visualizza dettagli -
2018 CIPM China | You have an invitation to the exhibition from LNEYA
2325The 56th (Fall 2018) National Pharmaceutical Machinery and China International Pharmaceutical Machinery Expo will be held from November 5th to November 7th, 2018 at Wuhan International Expo Center. As one of the professional communication platform...
Visualizza dettagli -
Quali sono le cause dei sedimenti nell'utilizzo del refrigeratore?
1766A chiller is a water-cooled unit that uses water circulation to provide a constant temperature, constant flow and constant pressure cooling water. The LNEYA cold water system has its own working principle, that is, a certain amount of water is inj...
Visualizza dettagli -
Quali sono i luoghi da scegliere per il freddo?
2276Oggi, con lo sviluppo della scienza e della tecnologia, molti clienti si concentrano sull'utilizzo di alte e basse temperature per migliorare la qualità dei prodotti e ridurne il costo. Quali sono quindi i punti a cui prestare attenzione quando si sceglie un modello? Uomo...
Visualizza dettagli
工業冷水機製造商 LNEYA Taiwan