LNEYA 高温・低温加熱式冷蔵庫 メンテナンス手順書
High and low temperature heating refrigerators use circulating water to replace circulating water regularly every month. If heat transfer oil is used, heat transfer oil should be replaced periodically every six months to one year. When the high and low temperature heating refrigerator is used for more than 150 degrees for a long time, check the heat transfer oil every three to six months, and decide whether to replace it according to the actual situation. When replacing the circulating oil, do not let water or foreign matter fall into the fuel tank.
Due to the thermal expansion and contraction during transportation and work in the middle of the machine, the nut of the terminal may be loosened, resulting in excessive current causing oxidation of the copper of the terminal, causing the short circuit to burn out the heating tube and the contactor. Maintenance of the heating tube of the high and low temperature heating chiller requires attention. Remove the sealing plate after the machine, open the wiring cover of the heating tube, and use the corresponding size of the sleeve for the nut of each connection and the nut at the joint of the copper piece (including under the copper piece). The nuts are all tightened and reinforced. If the water/oil scale of the high and low temperature heating refrigerator is serious, the whole heating pipe should be removed, and the surface of the heating pipe and the inside of the heating pipe should be cleaned and maintained to prevent the heating pipe from being burnt out due to poor circulation of the system.
When using the equipment, be sure to open the cooling water inlet and outlet valves on the back of the machine, with or without cooling water. The water circulation high and low temperature heating refrigerator may need to avoid the high pressure power failure of the machine due to thermal expansion during the heating process, and the pressure of the cooling water remaining inside the cooler may be prevented from being ruptured due to heat exchange expansion. Causes oil and water to mix and the machine does not work properly.
When not in use, be sure to disconnect the cooling water connection pipe, clean the remaining cooling water/heat transfer oil in the pipe, and store it in a dry place.
High and low temperature heating refrigerators need to be shut down for a period of time due to production. When restarting the machine, it is necessary to clean the filter and heating tube of the Y-type filter of the hot oil inlet to avoid the machine is not running for a long time, and the grease and impurities are deposited inside. The circulation is not smooth and the service life is shortened.
The maintenance of high and low temperature heating chillers is an important part of daily operation. Please keep in mind the high and low temperature heating chiller maintenance during the daily operation of the operator.
関連推奨品
-
研究室用冷却・加熱サーキュレーター選定ガイド
2019実験室用冷却・加熱サーキュレーター機器を購入する際、重要なポイントは生産要件を満たし、生産量を加速することである。そのため、機器の特性と技術的ソリューションを理解する必要があります。
詳細を見る -
LNEYAダイナミック温度制御システムの説明
2530The long-term research of LNEYA state temperature control system has been extended to the pharmaceutical chemical industry for a long time. The dynamic temperature control system is safe after-sales service, and it is strictly tested before leavin...
詳細を見る -
産業用高温・低温循環式チラーの冷却不能の解決策
2206When the high and low temperature recirculation chiller system is in use, what is the general cause if the high temperature cannot be reduced? How should we deal with such failures? First, let’s analyze the principle and method of cooling. The c...
詳細を見る -
TCU温度制御システムの原理
2563使用領域:様々な防爆領域で使用可能 温度制御ユニット(TCU)は主に精密化学、化学製薬、生物製薬プロセスの反応器加熱、冷却温度制御に使用される。それは自動温度制御を実現することができます。
詳細を見る
工業冷水機製造商 LNEYA Taiwan