冷凍ポンプ循環ポンプ故障の問題を解決するには?
Cooling heating devices are required in many chemical pharmaceutical and laboratory applications. Then how to solve the failure of the refrigeration heating device is also concerned about how to solve it?
1. Circulating pump failure
The circulating pump fault is generally the thermal overload of the circulating pump. There are four main reasons for this fault: The first is that the outer circulating pipeline of the heat conducting oil is too long. At this time, it is necessary to shorten the outer circulation pipeline, reduce the bending of the outer circulation pipeline, and then click Equipment electrical cabinet surface thermal relay Reset button; the second is the circulation pipeline blockage, cooling and heating device which is good to tell everyone, check the circulation pipeline cleaning, install the filter, and then click the equipment electrical cabinet surface thermal relay Reset button; The third is that the circulating pump voltage is too low. It is recommended to measure whether the circulating pump voltage meets the specified range, improve the voltage environment, click the Reset button of the equipment electrical cabinet surface thermal relay; the fourth is that the thermal relay regulates the current too small. At this time, it is recommended to measure the pump current. Compare the range of thermal relays, Phillips screwdriver to properly rotate the thermal relay current, click on the equipment electrical cabinet surface thermal relay Reset button.
2. Sensor failure
A great possibility of sensor failure is wiring error. Once the fault is encountered, check the wiring situation. At this time, you should pay attention to two situations, one is sensor short circuit or temperature <-160 °C, and the other is sensor open circuit or temperature > 349 ° C.
If the sensor is short-circuited or the temperature is <-160 °C, it may be the fault of the sensor itself. At this time, it is recommended to replace the sensor (there is a spare sensor in the equipment slot). If the sensor is wired incorrectly, It is recommended to check the wiring (corrected circuit diagram) correction.
If the sensor is open circuit or the temperature is >349°C, it should be the fault of the sensor itself. It is necessary to replace the sensor in time (there is a spare sensor in the equipment slot). At this time, it is necessary to pay attention to whether the temperature is too high and the temperature is too high. Check whether the temperature is exceeded. Range, if the equipment is not powered, pay attention to check the power supply of the equipment.
These two faults of the cooling heating device are still relatively simple. I believe that everyone who has seen the sharing can solve it very quickly.
関連推奨品
-
Application of supporting experimental equipment for refrigeration and heating system
2180The mixing tank with jacket and insulation layer is a three-layer insulation mixing tank with heating function, which has the functions of heating, insulation, mixing, etc. It is widely used in dairy, chemical, food, beverage, pharmaceutical, phar...
詳細を見る -
マイクロチャンネル反応器サーモスタット制御空冷式冷凍機コンプレッサーのトラブルシューティングメソッド
2086The air-cooled chiller is a temperature control device that is widely used in the microchannel reactor industry. Wuxi Guanya LNEYA has produced a cold and heat source thermostatically controlled air-cooled chiller for the microchannel reactor indu...
詳細を見る -
循環水冷却装置の冷媒は何ですか?
1924業界によって、循環水冷却装置に求められる冷凍温度は異なるため、選択される冷媒も異なります。循環式水冷装置を選ぶ際には、冷媒についての明確な理解も必要です。
詳細を見る -
工業用冷凍機が始動しない場合、以下の経験に基づいて除外することができる。
18901. Check the main circuit first. Such as whether there is electricity in the power supply, whether the voltage is normal, whether the fuse has been burned due to starting overload, whether the air switch trips, whether the switch contacts are good...
詳細を見る
工業冷水機製造商 LNEYA Taiwan