En heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Recomendaciones relacionadas
-
Prueba del motor del vehículo eléctrico Instrucciones del enfriador del sistema de refrigeración
2525Con el continuo desarrollo de los vehículos de nueva energía, también se está desarrollando el enfriador, el sistema de refrigeración para pruebas de motores eléctricos. El nuevo enfriador para pruebas de motores de vehículos de nueva energía puede ajustar de forma adaptativa el enfriador en función de la dinámica...
Ver detalles -
Characteristics of the evaporator of LNEYA screw chiller
2764The amount of cooling capacity of the screw chiller is directly related to the energy consumption and operational economics of the entire unit, which is worthy of attention. Especially when designing a cold station, there is no single chiller in t...
Ver detalles -
Mechanical cold assembly equipment is dedicated to cold processing assembly of metal parts
1796Mechanical cold assembly equipment is a method of processing workpieces at temperatures below -130℃. Cryogenic treatment can significantly improve the toughness of the workpiece without reducing the strength and hardness of the workpiece. Mechanic...
Ver detalles -
Heat transfer oil of explosion-proof cooling and heating circulator
1868Explosion proof cooling and heating circulator is a commonly used temperature control equipment in laboratories. It is used in pharmaceutical, chemical, biological and other industries to provide heat and cold sources for reactors, reactors, etc. ...
Ver detalles
工業冷水機製造商 LNEYA Taiwan