Failures of Heating Cooling Circulators
O circulador de aquecimento e arrefecimento can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Recomendações relacionadas
-
Rotary shaft freezing process in new energy automobile motor assembly
2387As the market for new energy vehicles has gradually expanded, the new energy motor manufacturing industry has also grown. Then, for the motor manufacturer, the rotor into the core process during the motor assembly process is particularly important...
Ver detalhes -
Sabe como poupar eletricidade no congelador a baixa temperatura?
1991The name of low temp freezer should be known to all, that is, the temperature inside the cabinet will be very low, so since the temperature is low, the relative power consumption is more. So under such circumstances, how can we save electricity? A...
Ver detalhes -
Sistema de controlo da temperatura de arrefecimento e aquecimento para apoio ao controlo da temperatura do reator de vidro
2116A indústria química do meu país desenvolveu-se rapidamente, e muitas empresas aceleraram o seu desenvolvimento estável. Na indústria química, o sistema de controlo da temperatura de aquecimento por refrigeração é combinado com um reator de vidro para controlar a temperatura...
Ver detalhes -
Celebrar o dia de Ano Novo e dar as boas-vindas ao Ano Novo
2287Quando os sinos ecoam no céu claro da noite,Quando o calendário vira para uma nova página,Estamos prestes a dar início a um novo ano.Obrigado pela colheita deste anoVamos ter uma experiência preciosa no nosso crescimentoHá sempre uma recompensa para o suorEste ano, nós...
Ver detalhes
工業冷水機製造商 LNEYA Taiwan