Como eliminar o alarme de alta pressão do congelador industrial de temperatura ultrabaixa?
During use, if a fault is not solved in time, it will affect the use of industrial ultra-low temperature freezer. How to solve the problem of high-pressure alarm?
If the water-cooled industrial ultra-low temperature freezer presents a high-pressure alarm, it is mainly related to the cooling water system. The water-cooled industrial ultra-low temperature freezer provides circulating cooling water through a cooling tower. If you forget to open the cooling water valve during use, If the cooling water cannot be circulated to the chiller, it will show high pressure problems. In addition, check whether the cooling water tower is defective and cannot safely and stably deliver the cooling water.
The cooling water flow is too small, or the temperature is too high and the heat dissipation is not good. This problem will also occur. If the cooling water flow is small, it is necessary to check whether the pipe size of the supporting device matches the size of the industrial ultra-low temperature freezer group, or whether the water valve can be opened to a large position; if it is an air-cooled industrial ultra-low temperature freezer, we need to look at the ambient temperature of the device If it can be exceeded, the heat must be exhausted in time to ensure that the industrial ultra-low temperature freezer unit is in an air circulating environment.
Industrial ultra-low temperature freezers are used for too long, and it is easy to accumulate dirt and other debris on the pipe wall or condenser copper pipes, which will inevitably affect the refrigeration cycle system. Therefore, we regularly provide industrial ultra-low temperature freezers according to the maintenance time recommended by the manufacturer Do maintenance cleaning work (cleaning the cooling water system, cleaning the condenser, changing the filter, etc.).
Check whether the refrigerant is excessive. The industrial ultra-low temperature freezer unit will be filled with the appropriate amount of refrigerant before leaving the factory. Therefore, if there is no leakage of the refrigerant, the user does not need to charge the refrigerant. Too much refrigerant will only affect the unit’s failure. use.
Whether the opening of the shrink valve is too small, for this, everyone can gradually adjust the opening of the shrink valve to eliminate the problem.
Recomendações relacionadas
-
Efeito de aplicação destrutiva de componentes no dispositivo de medição da temperatura de pastilhas
2028Os componentes podem ser testados em fontes frias de alta e baixa temperatura no dispositivo de medição de temperatura do chip microfluídico. A análise destrutiva de componentes é também uma aplicação importante, que pode medir a conceção, a estrutura e a ...
Ver detalhes -
Application of Industrial Air-Cooled Chiller in Chemical Industry
2247Industrial air-cooled chillers are applicable to glass reactors, metal reactors and bioreactors for temperature control. In the field of chemistry and pharmacy, air-cooled chillers usually carry out necessary temperature control together with the ...
Ver detalhes -
Cooperação mundial contra a COVID-19, a LNEYA presta apoio dentro das suas possibilidades
2528Na fase atual de transmissão do vírus, é imperativo estudar medicamentos e vacinas eficazes contra o coronavírus. Como fabricante de tecnologia de controlo de temperatura de alta precisão, acreditamos que é da nossa responsabilidade apoiar o trabalho...
Ver detalhes -
Como resolver a escassez de arrefecimento no sistema de ciclo de arrefecimento de semicondutores
2092O sistema de ciclo de arrefecimento de semicondutores é um dispositivo para arrefecer a parte geradora de calor dos semicondutores no equipamento de refrigeração. Se a capacidade de arrefecimento do sistema de circulação de arrefecimento de semicondutores LNEYA for insuficiente durante o funcionamento, o...
Ver detalhes
工業冷水機製造商 LNEYA Taiwan