Häufige Fehler von Salzsprüh-Umweltkammern
The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.
Salt spray environmental chamber for sale
The salt spray environmental chamber cannot be heated
1. Heat the solid state relay, if there is no output, replace the solid state relay.
2. If the external dial has no output, replace the external dial.
3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.
4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.
5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.
The heating temperature is abnormal
1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.
2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.
3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.
4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.
The external dial doesn’t display
1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.
2. Check whether or not the power switch is damaged. If damaged, replace the power switch.
3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.
4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.
5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.
If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.
Verwandte Empfehlungen
-
API exhibition LNEYA is brilliant
2289On October 14, 2020, the 85th China International Pharmaceutical Raw Materials/Intermediates/Packaging/Equipment Fair (API China for short) was grandly opened at Nanjing International Expo Center. The exhibition site was bustling and popular. LNEY...
Details anzeigen -
Cooling water circulator for BMS testing of liquid-cooled battery packs
1850As the weather temperature gradually increases, the use of liquid-cooled battery pack BMS test cooling water circulators also gradually reaches its peak. Liquid-cooled battery pack BMS test cooling water circulators have rated power ranging from s...
Details anzeigen -
Cooling and heating system TES for integrated circuit wafer cooling method
2169Integrated circuit products have been developed for more than 50 years and are the most advanced high-tech industries. Integrated circuit or chip / chip. The complete chip manufacturing process includes chip design, wafer manufacturing, package ma...
Details anzeigen -
The selection and maintenance of ultra-low temperature chiller
1722Today, a customer came to inquire about our need for a small laboratory water chiller. The specific working condition is to reduce the materials in the 20L stainless steel reactor from normal temperature to minus 30 ℃ in more than one hour. This p...
Details anzeigen
工業冷水機製造商 LNEYA Taiwan