Failures of Heating Cooling Circulators
The heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
관련 추천
-
영하 30도에서 에틸렌 글리콜 초저온 냉각기 작동 시 주의사항
2111영하 30도 에틸렌 글리콜 초저온 냉각기는 제약 및 화학 산업에서 많이 사용되는 장치입니다. 사용시 작동 사양에주의를 기울여 작동해야합니다. 부적절하게 사용하면 ...
세부 정보 보기 -
Energy Conservation is Crucial for Developing Ultra Low Temperature Recirculating Chillers
3662With the continuous development of chemical, refrigeration and pharmaceutical industries, environmental pollution is put on the agenda earlier. As an editor of which is the industrial chiller, I can say that except for corresponding to the environ...
세부 정보 보기 -
저온 챔버의 중요성
1828저온 챔버는 일반적으로 냉장고와 테스트 부품을 분리하고 테스트 부품을 룸 타입으로 만들어 직원이 작업 할 수 있도록하는 대형 극저온 또는 극저온 테스트 장치입니다. 이러한 테스트 룸을 저온 ...
세부 정보 보기 -
글리콜 냉동 시스템의 기술적 특성
1634At present, it is often necessary to provide process chilled water of 1-4℃ for the production line in the milk, beverage and other industries. Because the temperature of frozen water in 1-4℃ process is close to the freezing point of water at 0℃, i...
세부 정보 보기
工業冷水機製造商 LNEYA Taiwan