Cooling heating devices are required in many chemical pharmaceutical and laboratory applications. Then how to solve the failure of the refrigeration heating device is also concerned about how to solve it?
1. Circulating pump failure
The circulating pump fault is generally the thermal overload of the circulating pump. There are four main reasons for this fault: The first is that the outer circulating pipeline of the heat conducting oil is too long. At this time, it is necessary to shorten the outer circulation pipeline, reduce the bending of the outer circulation pipeline, and then click Equipment electrical cabinet surface thermal relay Reset button; the second is the circulation pipeline blockage, cooling and heating device which is good to tell everyone, check the circulation pipeline cleaning, install the filter, and then click the equipment electrical cabinet surface thermal relay Reset button; The third is that the circulating pump voltage is too low. It is recommended to measure whether the circulating pump voltage meets the specified range, improve the voltage environment, click the Reset button of the equipment electrical cabinet surface thermal relay; the fourth is that the thermal relay regulates the current too small. At this time, it is recommended to measure the pump current. Compare the range of thermal relays, Phillips screwdriver to properly rotate the thermal relay current, click on the equipment electrical cabinet surface thermal relay Reset button.
2. Sensor failure
A great possibility of sensor failure is wiring error. Once the fault is encountered, check the wiring situation. At this time, you should pay attention to two situations, one is sensor short circuit or temperature <-160 °C, and the other is sensor open circuit or temperature > 349 ° C.
If the sensor is short-circuited or the temperature is <-160 °C, it may be the fault of the sensor itself. At this time, it is recommended to replace the sensor (there is a spare sensor in the equipment slot). If the sensor is wired incorrectly, It is recommended to check the wiring (corrected circuit diagram) correction.
If the sensor is open circuit or the temperature is >349°C, it should be the fault of the sensor itself. It is necessary to replace the sensor in time (there is a spare sensor in the equipment slot). At this time, it is necessary to pay attention to whether the temperature is too high and the temperature is too high. Check whether the temperature is exceeded. Range, if the equipment is not powered, pay attention to check the power supply of the equipment.
These two faults of the cooling heating device are still relatively simple. I believe that everyone who has seen the sharing can solve it very quickly.
관련 추천
-
열전달 오일 순환 시스템의 냉각 성능 저하 문제를 해결하는 방법은 무엇입니까?
2067일부 열 전도성 오일 순환 시스템은 3개월 사용 후 이전만큼 좋지 않을 수 있습니다. 이때 전문 애프터 서비스 부서에 연락하여 사용자의 전원 공급 장치와 장비가 정상인지 확인하는 것이 좋습니다.
세부 정보 보기 -
스크류 워터 칠러의 장단점 분석
2382다른 유형의 냉각기에 비해 고품질 스크류 워터 칠러의 전체 에너지 소비량은 상대적으로 적습니다. 스크류 워터 칠러의 생산 기술이 더 발전하고 에너지가 덜 사용되기 때문에 장비는 ...
세부 정보 보기 -
Trouble shooting method for refrigeration device of air-cooled cold water mechanism chemical reactio
2203If air enters into the air-cooled cold water mechanism refrigeration unit system of the chemical reaction kettle, the impact will be many. Enter the air circulation system first, the pressure will rise, the heat dissipation effect of the condenser...
세부 정보 보기 -
TCU 온도 제어 모델 선택 및 기술 파라미터 설명
2611TCU 온도 제어는 2사이클 단위 단일 유체 온도 제어 시스템이기도 합니다. 모델을 선택할 때 전체 온도 제어 반응에 매우 중요합니다. 적절한 반응 시스템만이 발열 및 흡수에 사용됩니다.
세부 정보 보기
工業冷水機製造商 LNEYA Taiwan