Guidelines for Safe Use and Maintenance of Ultra-Low Temperature Freezer
Cryopreservation of samples is very important in modern biological research.
Correct use of cryogenic freezers (hereinafter referred to as refrigerators):
1. It is necessary to understand the influence of environment on refrigerators.
2. Familiar with basic operation, maintenance and maintenance knowledge
3. Develop good habits
Let the refrigerator breathe freely;
Let the air around the refrigerator circulate freely.
Free at least 15 cm of space around the refrigerator and 30 cm of space at the top.
Items should not be stacked at the top of the refrigerator to avoid affecting the heat dissipation at the top.
Give the refrigerator a reliable environment
Try to keep the ambient temperature around the refrigerator between 15 and 32 degrees Celsius for a long time.
Consider the difference of energy consumption between winter and summer, and control the environmental temperature reasonably.
If the ambient temperature is too high for a long time, the compressor life will be damaged.
Refrigerator heat dissipation will lead to higher ambient temperature
Eliminate the influence factors of heat exchange in refrigerators:
Don’t put the refrigerator near the heat source
Don’t put the refrigerator in the vent
Reserve special circuit for refrigerators
Basic operation
Turn on: After plugging in the power plug, press the switch in the lower right corner of the back, “O” stands for turning off, and “I” stands for turning on.
If the equipment is inclined more than 45 degrees when it is moved, the equipment will be placed vertically for 24 hours before start-up!!!
Open the battery switch: Open the white door under the front and press the switch down to “I”
Battery function: When the power is off, it can supply power to the display board and provide sound and light alarm.
We are a ultra low temperature freezer manufacturer. Welcome to choose our products.
Related recommendations
-
Hot and cold rapid temperature change test chamber common fault
2016In the high-temperature test of the hot and cold rapid temperature change test chamber, once the temperature change does not reach the test temperature value, in this case, the electrical system can be inspected and the faults can be eliminated on...
View details -
Common Faults of Salt Spray Environmental Chambers
1993The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault t...
View details -
Micro-channel reactor thermostatically controlled air-cooled chiller compressor troubleshooting meth
2086The air-cooled chiller is a temperature control device that is widely used in the microchannel reactor industry. Wuxi Guanya LNEYA has produced a cold and heat source thermostatically controlled air-cooled chiller for the microchannel reactor indu...
View details -
Rotary shaft freezing process in new energy automobile motor assembly
2387As the market for new energy vehicles has gradually expanded, the new energy motor manufacturing industry has also grown. Then, for the motor manufacturer, the rotor into the core process during the motor assembly process is particularly important...
View details
工業冷水機製造商 LNEYA Taiwan