What if the cooling rate of the cooling and heating temperature control system of the reactor slows
LNEYA cooling and heating temperature control system is a laboratory instrument and equipment for temperature control of supporting reactors in pharmaceutical and chemical laboratories and other industries. It adopts a fully closed circulation pipeline system to maintain the stability of temperature control of reactors in reactors.
The temperature range of the cooling and heating temperature control system of the reaction kettle is from – 120 ℃ to 350 ℃. Generally, the temperature selection and temperature setting will be carried out according to the needs of users. Within the specified time range, the cooling and heating temperature control system of the reaction kettle will reach the set temperature. However, if the cooling and heating temperature control system of the reaction kettle operates for a long time or operates improperly, the cooling of the cooling and heating temperature control system of the reaction kettle will slow down.
When the temperature drop of the reactor cooling and heating temperature control system is slow, it is necessary to touch the exhaust and suction pipes of the compressor of the main unit by hand. Once it is found that the temperature of the exhaust pipe of the new reactor cooling and heating temperature control system is not high, and the temperature of the suction pipe is not low but not frosted, it may be that the reactor cooling and heating temperature control system lacks refrigerant, Just add the refrigerant according to the steps of adding the refrigerant.
If the cooling rate of the reactor cooling heating temperature control system is slow, if it is not an electrical fault, it may be a problem of the refrigeration system. It is necessary to check the low temperature, exhaust and suction pressure of the refrigeration compressor in the refrigeration system of the new reactor cooling heating temperature control system. This may be due to insufficient refrigerant of the reactor cooling heating temperature control system.
LNEYA reminded that after finding out the slow fault of the cooling and heating temperature control system of the reaction kettle, it is necessary to solve the specific fault to better operate the new cooling and heating temperature control system of the reaction kettle.
Related recommendations
-
What are the temperature control performance of laboratory recirculating chillers
2065The laboratory recirculating chiller is a typical cooling circulator with heating function to ensure the repeatability of laboratory results. There are water-cooled or air-cooled refrigerators, which can be placed in a laboratory fume hood or set ...
View details -
Close contact between new energy vehicles and LNEYA battery test temperature control units
2689Based on the current situation of energy issues and the booming development of the automotive industry, new energy vehicles are as timely as the modern automotive industry. The battery part of the new energy vehicle plays a decisive role in the pe...
View details -
Fault Hint of Cryogenic Chiller Refrigeration System
1766When the high and low pressure of the system are higher than normal, there is usually air or excessive refrigerant in the system. At this time, we need to re-vacuum, and then add a proper amount of refrigerant to solve this problem. But there are ...
View details -
The Phenomenon of Ice Jam in Low Temperature Chiller System
1602The phenomenon of ice jam in low temperature chiller system is mainly due to the excessive water in the refrigeration system. With the continuous circulation of refrigerants, the water in the refrigeration system gradually concentrates at the outl...
View details
工業冷水機製造商 LNEYA Taiwan