Several Common Troubleshooting Methods for Ultra-low Temperature Freezer in Laboratories
Although the refrigeration system and control circuit of ultra low temperature freezer in laboratory are complex, the common faults are mainly circulating system and circuit part.
Following are several common faults of ultra-low temperature refrigerators and their troubleshooting methods:
1. After power supply is switched on, the air switch of the main power supply is automatically disconnected.
Elimination method: Check whether the compressor of refrigeration system is damaged or not, and replace the parts if damaged. After replacing the fittings, turn on the machine to check whether it is normal. The fault is usually caused by unstable power supply, because the * stage starts first, so the damage is usually in stage. In this case, as long as the delay regulated power supply is installed, the fault will not occur again.
2. Stop working before the temperature in the box reaches the set temperature.
Exclusion: Although the temperature in the box has not reached the set temperature, the panel display temperature has reached the set temperature. This is due to the damage of the thermistor which detects the temperature and the need to replace the parts. If there is no original accessories, a set of temperature control system can be re-installed to eliminate the failure.
3. Poor refrigeration effect may be refrigerant leakage
4. If the ultra-low temperature refrigerator does not start in winter, it may be that the thermostat regulation value is too small.
We are a ultra low temperature freezer manufacturer. Welcome to choose our products.
Related recommendations
-
Guidelines for Safe Use and Maintenance of Ultra-Low Temperature Freezer
19061. It is necessary to understand the influence of environment on refrigerators. 2. Familiar with basic operation, maintenance and maintenance knowledge 3. Develop good habits Let the refrigerator breathe freely; Let the air around the refrigerator...
View details -
Meet at booth 6002 of Shanghai World Expo Hall【Testing Expo】9.14-9.16
259Exhibition name: Shanghai International Automobile Testing and Quality Control ExhibitionExhibition time: September 14-16, 2020Booth number: 6002Exhibition address: Shanghai World Expo Exhibition Hall, Hall 1
View details -
Method for controlling shutdown of water cooled chiller of single reaction kettle
2336The reaction kettle water cooled chiller has a wide range of accessories and equipment. In actual use, it needs to be provided with regular rest time. In order to maintain the stable performance of the refrigerator during long-term operation, it i...
View details -
The Advantage Parameter Explanation Of The Walk-In Lab Water Chiller
1736In the experimental equipment, the walk-in lab water chiller is used for all kinds of electronic components in high and low temperature or hot and wet environment to test the requirements of various performance indicators. It is generally applicab...
View details
工業冷水機製造商 LNEYA Taiwan